Institutional Repository of Chinese Acad Sci, Inst Automat, CAS Key Lab Mol Imaging, Beijing 100190, Peoples R China
In vivo quantitative bioluminescence tomography using heterogeneous and homogeneous mouse models | |
Liu, Junting1; Wang, Yabin2; Qu, Xiaochao1; Li, Xiangsi1; Ma, Xiaopeng1; Han, Runqiang1; Hu, Zhenhua1; Chen, Xueli1; Sun, Dongdong2; Zhang, Rongqing2; Chen, Duofang1; Chen, Dan1; Chen, Xiaoyuan4; Liang, Jimin1; Cao, Feng2; Tian, Jie1,3; J. Liang, F. Cao, J. Tian | |
发表期刊 | OPTICS EXPRESS |
2010-06-07 | |
卷号 | 18期号:12页码:13102-13113 |
文章类型 | Article |
摘要 | Bioluminescence tomography (BLT) is a new optical molecular imaging modality, which can monitor both physiological and pathological processes by using bioluminescent light-emitting probes in small living animal. Especially, this technology possesses great potential in drug development, early detection, and therapy monitoring in preclinical settings. In the present study, we developed a dual modality BLT prototype system with Micro-computed tomography (MicroCT) registration approach, and improved the quantitative reconstruction algorithm based on adaptive hp finite element method (hp-FEM). Detailed comparisons of source reconstruction between the heterogeneous and homogeneous mouse models were performed. The models include mice with implanted luminescence source and tumor-bearing mice with firefly luciferase report gene. Our data suggest that the reconstruction based on heterogeneous mouse model is more accurate in localization and quantification than the homogeneous mouse model with appropriate optical parameters and that BLT allows super-early tumor detection in vivo based on tomographic reconstruction of heterogeneous mouse model signal. (C) 2010 Optical Society of America |
关键词 | - |
WOS标题词 | Science & Technology ; Physical Sciences |
关键词[WOS] | SOURCE RECONSTRUCTION ; IMAGE-RECONSTRUCTION ; LIGHT ; PROPAGATION ; ALGORITHM ; TISSUE ; CELL |
收录类别 | SCI |
语种 | 英语 |
WOS研究方向 | Optics |
WOS类目 | Optics |
WOS记录号 | WOS:000278527700115 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.ia.ac.cn/handle/173211/3947 |
专题 | 中国科学院分子影像重点实验室 |
通讯作者 | J. Liang, F. Cao, J. Tian |
作者单位 | 1.Xidian Univ, Sch Life Sci & Technol, Life Sci Res Ctr, Xian 710071, Peoples R China 2.Fourth Mil Med Univ, Xijing Hosp, Dept Cardiol, Xian 710032, Peoples R China 3.Chinese Acad Sci, Inst Automat, Beijing 100190, Peoples R China 4.NIBIB, LOMIN, NIH, Bethesda, MD 20892 USA |
推荐引用方式 GB/T 7714 | Liu, Junting,Wang, Yabin,Qu, Xiaochao,et al. In vivo quantitative bioluminescence tomography using heterogeneous and homogeneous mouse models[J]. OPTICS EXPRESS,2010,18(12):13102-13113. |
APA | Liu, Junting.,Wang, Yabin.,Qu, Xiaochao.,Li, Xiangsi.,Ma, Xiaopeng.,...&J. Liang, F. Cao, J. Tian.(2010).In vivo quantitative bioluminescence tomography using heterogeneous and homogeneous mouse models.OPTICS EXPRESS,18(12),13102-13113. |
MLA | Liu, Junting,et al."In vivo quantitative bioluminescence tomography using heterogeneous and homogeneous mouse models".OPTICS EXPRESS 18.12(2010):13102-13113. |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
In vivo quantitative(336KB) | 期刊论文 | 作者接受稿 | 开放获取 | CC BY-NC-SA | 浏览 下载 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。